Method for express-calibration of the sensitivity of a multichannel adaptive phase demodulator.
Keywords:
adaptive phase demodulator, calibration, dynamic hologram, photorefractive crystalAbstract
In this paper, we present a method for calibrating a multichannel adaptive phase demodulator. The proposed method consists of introducing a known phase modulation with variable amplitude into the channels of the demodulator, constructing a transient response and determining with it the calibration coefficient for each channel of the demodulator. The process of calibration is automated and takes 2 seconds. Express calibration method gives an opportunity to mitigate a sensitivity spread through the channels of adaptive phase demodulator.
Downloads
Published
2019-05-13
How to Cite
Безрук, М. Н., Ромашко, Р. В., Кульчин, Ю. Н., Ермолаев, С. А., & Ноткин, Б. С. (2019). Method for express-calibration of the sensitivity of a multichannel adaptive phase demodulator. Vestnik of the Far East Branch of the Russian Academy of Sciences, (2), 117–120. Retrieved from http://vestnikdvo.ru/index.php/vestnikdvo/article/view/198